Calibration Test Blood Pressure For Baumanometer Instruments Kit Black Each
- MFR:
- 2941
- Medex SKU:
- WAB-2941
- Packing Info:
- 1/Each
- Usually Ships:
- 5 - 7 Business Days
- Freight Quote Button:
- False
Description
TESTING INSTRUMENT ACCURACY
Your goal is to test instrument accuracy, not the process of blood pressure measurement. You will need a Baumanometer® instrument (mercury-gravity standard) and a “Y” connector with an inflation bulb and valve attached (see below for Test Kit ordering information). Connect one leg to the Baumanometer® instrument and the other to the blood pressure instrument to be tested.
Note: Cuffs and Bags are not used when conducting this test.
The Pressure Standard: A Baumanometer® instrument is functioning correctly and should be used as the pressure standard if:
The mercury meniscus is at zero with no pressure applied to the instrument.
The instrument is in a vertical position. note: The Standby model is calibrated at a 20o angle
The instrument responds promptly to pressure changes.
Any two Baumanometer® instruments, regardless of age, will provide accurate linear pressure readings at every pressure level if they meet the stated criteria for a correctly functioning manometer.
Test Procedure: Check each instrument to be sure that it is at zero. Slowly inflate the instruments to 250 mmHg and compare the readings. They should be the same, however a deviation of +/- 3 mmHg is acceptable. Repeat this procedure at 200 mmHg, 150 mmHg, 100 mmHg, 50 mmHg and 0 mmHg. If the deviation is greater than +/- 3 mmHg at any of these points, the instrument being tested is inaccurate and needs adjustment or repair.
Calibration "Y" connector (To use your Baumanometer® instrument as a standard for pressure measurement).