Skip to main content
(0) Write a Review

On-Q Pump with On-Demand 400 mL, CB003

$3,588.95
For larger quantities: Request a Quote
Medex SKU:
AVA-CB003
Packing Info:
5/Case
Usually Ships:
7 - 10 Business Days

Description

The CB003 is an elastomeric infusion system that uses the mechanical tension of a pressurized internal balloon reservoir to drive medication through a flow-restricting capillary. It is a "fixed-flow" device, meaning the dosage rate is pre-set at the factory and cannot be manipulated by the patient, which enhances safety in home-care settings.

Technical Specifications

FeatureSpecification
Fill Volume400 ml
Flow Rate10 ml/hr (Fixed)
DurationApprox. 40 hours (~1.6 days)
Tubing TypeSingle Lead
NRFit™ ConnectorYes (Yellow-coded for regional anesthesia safety)

Key Features

  • High Flow Rate (10 ml/hr): This model is designed to deliver a higher volume of anesthetic per hour. This is particularly useful for saturating larger surgical wounds or providing a more "dense" nerve block.

  • NRFit™ Safety: Utilizes specialized connectors that are physically incompatible with Luer-lock IV connections, preventing the potentially fatal error of injecting local anesthetic into a patient's intravenous line.

  • Mechanical Reliability: Because it lacks electronics, motors, or batteries, it is completely silent and provides a consistent flow based on the physical properties of the elastomeric balloon.

  • Integrated Filtration: Contains a 0.2-micron air-venting filter to protect against air embolisms and a 1.2-micron particulate filter to ensure the purity of the medication delivered to the patient.

Clinical Application

The CB003 (400ml / 10ml/hr) is typically chosen for procedures where the peak pain is intense but expected to subside more quickly, or where the surgical site is broad, such as:

  • Large Abdominal Procedures: For surgeries like a ventral hernia repair where a larger volume of fluid is needed to cover the incision.

  • Thoracic Surgery: Used in paravertebral or intercostal blocks where a higher flow rate may improve the spread of the anesthetic.

  • Orthopedic Field Blocks: When a high-volume "soak" of a joint or muscle bed is required immediately following surgery.